Application of a Single-Crystal CVD Diamond Detector for Fast Neutron Measurement in High Dose and Mixed Radiation Fields

Makoto I. Kobayashi*, Sachiko Yoshihashi, Kunihiro Ogawa, Mitsutaka Isobe, Tsukasa Aso, Masanori Hara, Siriyaporn Sangaroon, Sachie Kusaka, Shingo Tamaki, Isao Murata, Sho Toyama, Misako Miwa, Shigeo Matsuyama, Masaki Osakabe

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

抄録

This article presents the method to evaluate the fast neutron energy spectrum using the single-crystal chemical vapor deposition (CVD) diamond detector to be applicable on the radiation monitoring in advanced scientific/engineering systems usually characterized with mixed and high-dose radiation field. The pulse shape discrimination (PSD) based on the shape and the width of a pulse was applied to extract events, in which fast neutron hits at the specific depth of the single-crystal diamond. Unfolding of the measured spectrum for extracted pulses could deduce the neutron energy spectrum. Experiments using monoenergetic neutron sources demonstrated the reliable capability of this method to evaluate the neutron energy spectrum quantitatively.

本文言語英語
論文番号6010808
ジャーナルIEEE Transactions on Instrumentation and Measurement
73
DOI
出版ステータス出版済み - 2024

ASJC Scopus 主題領域

  • 器械工学
  • 電子工学および電気工学

フィンガープリント

「Application of a Single-Crystal CVD Diamond Detector for Fast Neutron Measurement in High Dose and Mixed Radiation Fields」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル